FDG-PET/CT-based Gross Tumor Volume Contouring for Radiation Therapy Planning: An Experimental Phantom Study
スポンサーリンク
概要
- 論文の詳細を見る
As there is continuing controversy over the role of F-18-fluorodeoxyglucose (FDG)-positron emission tomography (PET)/CT-fused imaging in radiation therapy (RT) planning, we performed a phantom study to assess the feasibility of FDG-PET/CT-based gross tumor volume (GTV) contouring. The phantom set, consisting of an elliptical bowl and 6 spheres measuring from 10–37 mm in diameter, were filled with FDG to obtain 3 source-to-background ratios (SBRs) of 4, 8, and 16. The ratio to maximum intensity at 5% intervals was applied as the threshold for contouring. The ratio between contoured- and actual volumes (volume ratio) was calculated, and the threshold ratio was selected to provide a volume ratio close to 100%. To consider the clinical application, we applied the threshold value (maximum intensity × threshold ratio) for the largest 37-mm sphere to the 6 spheres. The threshold ratio and the volume ratio in 6 spheres with 3 SBRs were compared using the Friedman test. Threshold ratios ranged from 25–80%; they were higher for smaller spheres (p = 0.003) and lower SBRs (p < 0.001). The volume ratios with the threshold value for the largest 37-mm sphere were lower in smaller spheres (p = 0.010). These results suggest that smaller lesions and higher background activities require a higher threshold ratio and smaller lesions a lower threshold value. FDG-PET/CT-fused imaging should not be used as a single modality but rather to obtain supplemental information in RT planning. The contoured GTV should be adjusted based on clinical data including conventional images.
著者
-
Tomiguchi Seiji
Department Of Diagnostic Radiology School Of Health Sciences Kumamoto University
-
Shiraishi Shinya
Department Of Diagnostic Radiology Graduate School Of Medical Sciences Kumamoto University
-
Yoshida Morikatsu
Department Of Biochemistry National Cardiovascular Center Research Institute
-
Yamashita Yasuyuki
Department Of Diagnostic Imaging Faculty Of Medical And Pharmaceutical Sciences Kumamoto University
-
Oya Natsuo
Department Of Radiation Oncology And Image-applied Therapy Graduate School Of Medicine Kyoto Univers
-
Kawanaka Koichi
Department Of Diagnostic Imaging Faculty Of Medical And Pharmaceutical Sciences Kumamoto University
-
Tashiro Kuniyuki
Department Of Diagnostic Imaging Faculty Of Medical And Pharmaceutical Sciences Kumamoto University
-
Murakami Ryuji
Department Of Radiation Oncology Graduate School Of Medical Sciences Kumamoto University
-
NAKAGUCHI Yuji
Department of Radiological Technology, Kumamoto University Hospital
-
TSUJITA Naoko
Department of Medical Imaging, Faculty of Life Sciences, Kumamoto University
-
TOYA Ryo
Departments of Radiation Oncology, Faculty of Life Sciences, Kumamoto University
-
SAKAMOTO Fumi
Department of Diagnostic Radiology, Faculty of Life Sciences, Kumamoto University
-
OYA Natsuo
Departments of Radiation Oncology, Faculty of Life Sciences, Kumamoto University
-
KAWANAKA Koichi
Department of Diagnostic Radiology, Faculty of Life Sciences, Kumamoto University
-
MURAKAMI Ryuji
Department of Medical Imaging, Faculty of Life Sciences, Kumamoto University
-
MURAKAMI Ryuji
Departments of Medical Imaging, Faculty of Life Sciences, Kumamoto University
-
YOSHIDA Morikatsu
Department of Diagnostic Radiology, Faculty of Life Sciences, Kumamoto University
-
TSUJITA Naoko
Departments of Medical Imaging, Faculty of Life Sciences, Kumamoto University
-
SHIRAISHI Shinya
Department of Diagnostic Radiology, Faculty of Life Sciences, Kumamoto University
関連論文
- Contrast Enhancement for Whole-body Screening Using Multidetector Row Helical CT : Comparison between Uniphasic and Biphasic Injection Protocols
- PJ-052 Quantitative Assessment of Total Coronary Artery Plaques Volume Using Multi-slice Computed Tomography is Associated with Metabolic Syndrome(PJ009,CT/MRI (Coronary/Vascular) 4 (I),Poster Session (Japanese),The 73rd Annual Scientific Meeting of The J
- PJ-052 Quantitative Assessment of Total Coronary Artery Plaques Volume Using Multi-slice Computed Tomography is Associated with Metabolic Syndrome(PJ009,CT/MRI (Coronary/Vascular) 4 (I),Poster Session (Japanese),The 73rd Annual Scientific Meeting of The J
- 7 Clinical Value of a Novel Cardiac Image Fusion Software CardIQ for Three-dimensional Integration of SPECT and 64-slice CT Coronary Angiography(Frontiers in Coronary Imaging,Symposium 15 (SY-15) (A),The 73rd Annual Scientific Meeting of The Japanese Circ
- Role of cardiac computed tomography in patients with suspected coronary artery disease : interaction with nuclear cardiology
- Initial experience with X-ray CT based attenuation correction in myocardial perfusion SPECT imaging using a combined SPECT/CT system
- Attenuation correction using combination of a parallel hole collimator and an uncollimated non-uniform line array source
- Attenuation correction using asymmetric fanbeam transmission CT on two-head SPECT system
- CT Portography by Multidetector Helical CT : Comparison of Three Rendering Models
- Ghrelin Suppresses Cardiac Sympathetic Activity and Attenuates Early Left Ventricular Remodeling in Rats with Myocardial Infarction(Old Myocardial Infarction/Remodeling 1 (IHD), The 69th Annual Scientific Meeting of the Japanese Circulation Society)
- A Case of Metachronous Bilateral Breast Cancer with Bilateral Radiation Pneumonitis After Breast-conserving Therapy
- Influence of Radiation Dose Rate and Lung Dose on Interstitial Pneumonitis after Fractionated Total Body Irradiation : Acute Parotitis May Predict Interstitial Pneumonitis
- Radiation Therapy for Brain Metastases from Breast Cancer
- Late Skin and Subcutaneous Soft Tissue Changes after 10-Gy Boost for Breast Conserving Therapy
- A Case of Allergic Reaction to Surgical Metal Clips Inserted for Postoperative Boost Irradiation in a Patient Undergoing Breast-Conserving Therapy
- Meningeal Carcinomatosis in Patients with Breast Cancer: Report of 8 Patients
- Clinical Results of Breast Conserving Therapy for Stage I and II Breast Cancers : Assessment of Local Recurrences in Relation to Surgical Margins
- Comparison of in vivo Efficacy of Hypoxic Cytotoxin Tirapazamine and Hypoxic Cell Radiosensitizer KU-2285 in Combination with Single and Fractionated Irradiation
- Identification and characterization of focal ground-glass opacity in the lungs by high-resolution CT using thin-section multidetector helical CT : experimental study using a chest CT phantom
- Prediction of aortic peak enhancement in monophasic contrast injection protocols at multidetector CT : phantom and patient studies
- Digital Cine Angiography Permits Radiation Dose Reduction without Reduction in Image Quality
- OJ-200 Increased Pericardial Fat Volume is Associated with Presence of Coronary Artery Plaques with or without Coronary Stenosis and Positive Remodeling(OJ34,Atherosclerosis (Clinical/Pathophysiology) (IHD),Oral Presentation (Japanese),The 73rd Annual Sci
- Azelnidipine, a long-acting calcium channel blocker, could control hypertension without decreasing cerebral blood flow in post-ischemic stroke patients. A ^I-IMP SPECT follow-up study
- The diagnostic efficacy of FDG-PET in the local recurrence of hilar bile duct cancer
- Clinical Role of Non-Contrast Magnetic Resonance Angiography for Evaluation of Renal Artery Stenosis
- Management of sentinel lymph nodes in malignant skin tumors using dynamic lymphoscintigraphy and the single-photon-emission computed tomography/computed tomography combined system
- Development of a collimator blurring compensation method using fine angular sampling projection data in SPECT
- Optimum energy window setting on Hg-201 x-rays photopeak for effective Tl-201 imaging
- Accurate scatter correction for transmission computed tomography using an uncollimated line array source
- Hyperoxia-induced Acute Lung Injury Using a Pig Model: Correlation between MR Imaging and Histologic Results
- Phantom Evaluation of Scatter and Attenuation Correction in Thallium-201/Technetium-99m Acquisition in Myocardial Perfusion Single-photon Emission Computed Tomography
- Diagnostic accuracy of simultaneous acquisition of transmission and emission data with technetium-99m transmission source thallium-201 myocardial SPECT
- Evaluation of persistence of ductus venosus with Tc-99m DTPA galactosyl human serum albumin liver scintigraphy and I-123 iodoamphetamine per-rectal portal scintigraphy
- A new method for crosstalk correction in simultaneous dual-isotope myocardial imaging with Tl-201 and I-123
- Radiation dose reduction in hepatic multidetector computed tomography with a novel adaptive noise reduction filter
- Dynamic MR Imaging and Radiotherapy
- Reduction of Radiation Dose at HRCT of the Temporal Bone in Children
- Ectopic cervical thymus in an infant
- Optimal dose and injection duration (injection rate) of contrast material for depiction of hypervascular hepatocellular carcinomas by multidetector CT
- Multidetector CT findings suggesting a perforation site in the gastrointestinal tract : analysis in surgically confirmed 155 patients
- Differential Diagnosis of Small Round Cervical Lymph Nodes: Comparison of Power Doppler US with Contrast-enhanced CT and Pathologic Results
- Time Course of Reoxygenation in Experimental Murine Tumors after Carbon-beam and X-ray Irradiation
- Detection of Hepatocellular Carcinoma Using Double-echo FLASH Sequence during the Hepatic Arterial Phase
- Role f Preoperative Helical CT before Laparoscopic Cholecystectomy : Evaluation of Gallbladder and Peritoneal Adhesion
- Evaluation of Hypervascular Hepatocellular Carcinoma in Cirrhotic Liver by Means of Helical CT : Comparison of Different Contrast Medium Concentrations within the Same Patient
- MR Imaging of the Anterior Cruciate Ligament: Value of Thin Slice Direct Oblique Coronal Technique
- Percutaneous Microwave Coagulation Therapy: An Experimental Study
- Metastatic Liver Tumor from Cystic Ovarian Carcinomas:CT and MRI Appearance
- The Value of Fast and Ultrafast T2-weighted MR Imaging Sequences in Hepatic Enhancement with Ferumoxides:Comparison with Conventional Spin-echo Sequence
- Calculation of T2 Relaxation Time from Ultrafast Single Shot Sequences for Differentiation of Liver Tumors:Comparison of Echo-planar, HASTE, and Spin-echo Sequences
- Predictive Factors for Lung Dose Reduction by Respiratory Gating at Radiotherapy for Lung Cancer
- Triplet Sensitized Laser Flash Photolysis of BNN3, Caged Nitric Oxide
- Quantitative Evaluation of Measurement Accuracy for Three-dimensional Angiography System Using Various Phantoms
- Intraarterial Infusion Chemotherapy for Head and Neck Cancers : Evaluation of Tumor Perfusion with Intraarterial CT during Carotid Arteriography
- Usefulness of Source Images from Three-dimensional Time-of-flight MR Angiography after Treatment of Cavernous Dural Arteriovenous Fistulas
- Totally Occluded Iliac Arteries: Long-term Results of Pecutaneous Transluminal Angioplasty
- Endometrial adenocarcinoma following the conservative treatment of an atypical polypoid adenomyoma
- Intracranial Artery Dissections : Serial Evaluation with MR Imaging, MR Angiography, and Source Images of MR Angiography
- The detection rates and tumor clinical/pathological stages of whole-body FDG-PET cancer screening
- Synchronous spontaneous arteriovenous fistula and aneurysm of the deep femoral artery incidentally found in a patient with aortic aneurysmal dissection
- PREPARATION AND EVALUATION OF W/O TYPE AND W/O/W TYPE EMULSIONS CONTAINING AN ANTICANCER DRUG
- Enhanced Cell Killing by Overexpression of Dominant-negative Phosphatidylinositol 3-Kinase Subunit, ΔAp85, Following Genotoxic Stresses
- 5-Fluorouracil Intra-arterial Infusion Combined With Systemic Gemcitabine for Unresectable Pancreatic Cancer
- Preoperative Detection of Liver Metastases Secondary to Pancreatic Cancer : Utility of Combined Helical Computed Tomography During Arterial Portography With Biphasic Computed Tomography-Assisted Hepatic Arteriography
- Measurement of Effective Renal Plasma Flow (ERPF) with 123I - Orthoiodohippurate (I - 123 - OIH)
- Quantitative evaluation of the regional hepatic reserve by ^Tc-GSA dynamic SPECT before and after chemolipiodolization in patients with hepatocellular carcinoma
- Azelastine : Its Clinical Application for Radiation Dermatitis
- Quantitative renography with the organ volume method and interporative background subtraction technique
- Transjugular Intrahepatic Portosystemic Shunt: Initial Clinical Experience and Three-Year Follow-up
- Functional Evaluation of Transplanted Kidneys by Gd-DTPA Enhanced Turbo FLASH MR Imaging
- A Segmentectomy of the Right Upper Lobe Has an Advantage over a Right Upper Lobectomy Regarding the Preservation of the Functional Volume of the Right Middle Lobe : Analysis by Perfusion Single-Photon Emission Computed Tomography/Computed Tomography
- Chemoradiotherapy for pancreatic cancer : current status and perspectives
- MR Imaging of Hepatocellular Carcinoma Arising in Genetic Hemochromatosis
- Detection of liver metastases secondary to pancreatic cancer: Utility of combined helical computed tomography during arterial portography with biphasic computed tomography-assisted hepatic arteriography
- Estimation of 123I-IMP arterial blood activity from dynamic planar imaging of the chest using a graph plot method for the quantification of regional cerebral blood flow
- Hypofractionated stereotactic radiotherapy for acoustic neuromas: safety and effectiveness over 8 years of experience.
- Clinical usefulness of a collimator distance dependent resolution recovery in myocardial perfusion SPECT : a clinical report from a single institute
- Reoxygenation of Experimental Murine Tumors after Carbon Beam Irradiation : Evaluation by Colony Assay.
- Does adding FDG-PET to MRI improve the differentiation between primary cerebral lymphoma and glioblastoma? Observer performance study
- Liver functional volumetry for portal vein embolization using a newly developed ^Tc-galactosyl human serum albumin scintigraphy SPECT-computed tomography fusion system
- Type 1 Perimedullary Arteriovenous Fistula with Subarachnoid Hemorrhage : Utility of Contrast-enhanced 3D Gradient-echo Technique
- Multi-tensor Tractography of the Motor Pathway at 3T : A Volunteer Study
- ^C MR Imaging of Methionine-rich Gliomas at 4.7T : A Pilot Study
- Repeat FDG-PET for Predicting Pathological Tumor Response and Prognosis after Neoadjuvant Treatment in Nonsmall Cell Lung Cancer : Comparison with Computed Tomography
- Improvement of the ^Tc-ECD brain uptake ratio (BUR) method for measurement of cerebral blood flow
- FDG-PET/CT-based Gross Tumor Volume Contouring for Radiation Therapy Planning: An Experimental Phantom Study
- Histopathological Changes in Parotid and Submandibular Glands of Patients Treated with Preoperative Chemoradiation Therapy for Oral Cancer
- Evaluation of Deep Vein Thrombosis With Reduced Radiation and Contrast Material Dose at Computed Tomography Venography : Clinical Application of a Combined Iterative Reconstruction and Low-Tube-Voltage Technique
- A Case of Cortisol Producing Adrenal Adenoma Associated with a Latent Aldosteronoma: Usefulness of the ACTH Loading Test for the Detection of Covert Aldosteronism in Overt Cushing Syndrome
- Fast neutron spectra transmitted through iron and sodium slabs.
- Guidelines on the Use of Iodinated Contrast Media in Patients With Kidney Disease 2012 : Digest Version
- Accumulation of pericardial fat correlates with left ventricular diastolic dysfunction in patients with normal ejection fraction
- Guidelines on the use of iodinated contrast media in patients with kidney disease 2012: digest version