コバスTaqMan HBV「オート」v2.0における同一時の血清検体と血漿検体のHBV DNA検出率の検討
スポンサーリンク
概要
- 論文の詳細を見る
The sensitivity in serum and plasma for HBV DNA was evaluated by using 104 clinical specimens from 52 patients who were treated with entecavir for ≥1 year and continued ALT levels ≤30 IU/l. The measurement employed the COBAS TaqMan HBV v2.0. Twenty-five specimens (24.0%) were detected from both serum and plasma, and 41 specimens (39.4%) were not detected from both. On the other hand, there were 32 specimens (30.8%) with detectable from plasma but undetectable from serum, and only 6 specimens (5.8%) with detectable from serum but undetectable from plasma. This result suggested the sensitivity of HBV DNA using plasma specimen is more sensitive than that of serum specimen with statistical significance (p<0.001).