Multivaneにおける画像特性の基礎的検討
スポンサーリンク
概要
- 論文の詳細を見る
Deterioration in the image because of a patient’s movement always becomes a problem in the MRI inspection. To solve this problem, the imaging procedure named Multivane was developed. The principle is similar to the periodically rotated overlapping parallel lines with enhanced reconstruction (PROPELLER) method. As for Multivane, the effect of the body motion correction is high. However, the filling method of k space is different than a past Cartesian method. A basic examination of the image characteristic of Multivane and Cartesian was utilized along with geostationary phantom. The examination items are SNR, CNR, and a spatial resolution. As a result, Multivane of SNR was high. Cartesian of the contrast and the spatial resolution was also high. It is important to recognize these features and to use Multivane.