Evaluation of Mechanical Seals for Rotary Blood Pumps by a Chronic Animal Experiment
スポンサーリンク
概要
- 論文の詳細を見る
The integrity of the shaft seal is a critical factor in the success of an implantable rotary blood pump. A precision mechanical seal has been developed to address this requirement. The seal incorporates a recirculating cooling water system to provide lubrication and prevent deposition of blood elements. The objective of this study is to evaluate the sealing properties of this mechanical seal by a chronic animal experiment. The leakage rate of plasma into the cooling water through the mechanical seal was measured by monitoring the infiltration of sodium ions from bovine blood. The leakage rate of the cooling water into the blood through the mechanical seal was measured by monitoring the volume of the cooling water in the reservoir. The cooling water temperatures were monitored at 2 points to evaluate heat dissipation. Over a period of 153 days, the leakage rate of plasma was 0.022ml/day and the leakage rate of the cooling water was 0.497 ml/day. The heat dissipation by the cooling water was 10-14 W in the range of flow rate at 50-100 ml/min. In conclusion, the mechanical seal performed acceptably well to support its use in a rotary blood pump for chronic application.
- ライフサポート学会の論文
- 2003-04-15
著者
-
Tomioka Jun
Department Of Mechanical Engineering School Of Science And Enginedring Waseda University
-
Tomioka Jun
Department Of Mechanical Engineering Waseda University
-
Yamazaki Kenji
Department Of Cardiovascular Surgery The Heart Institute Of Japan Tokyo Women's Medical Univers
-
AKIMOTO Takehide
University of Pittsburgh
-
Yamazaki Shun'ichi
Sun Medical Technology Research Corporation
-
Yamazaki Kenji
Tokyo Women's Medical College
-
Yamazaki Kenji
The Department Of Cardiovascular Surgery The Heart Institute Of Japan Tokyo Women's Medical Uni
関連論文
- 2 Next Generation LVAS as a Feasible Alternative to Heart Transplantation(Ventricular Assist Device: An Update,Topic 6 (TP-6) (IHD),The 73rd Annual Scientific Meeting of The Japanese Circulation Society)
- 3 Potential for Destination Therapy with Next Generation Left Ventricular Assist System (LVAS)(Impacts of latest developments of artificial organs/systems in the treatment of cardiovascular diseases,Topic 5 (TP-5) (IHD),Special Program,The 72nd Annual Sci
- 7 Next Generation LVAS as a Feasible Alternative to Heart Transplantation(Therapeutic Strategy for Severe Heart Failure-From LVAD to Heart Transplantation-,Symposium 11 (SY-11) (H),The 73rd Annual Scientific Meeting of The Japanese Circulation Society)
- 4 Clinical Trial of EVAHEART : Next Generation Left Ventricular Assist Device(Advance in Medical and Biological Engineering,Topic 1 (TP1) (I),The 70th Anniversary Annual Scientific Meeting of the Japanese Circulation Society)
- 6 Japanese Clinical Trial Results of Next Generation LVAS "EVAHEART"(Symposium 8 (SY-08) (M) New Strategies for Management of Severe Heart Failure with New Devices and Surgery,Special Program,The 72nd Annual Scientific Meeting of the Japanese Circulation
- Left Ventricular Mass Index Reduction Early after an Isolated Aortic Valve Replacement with St. Jude Medical 19A-HP
- Mechanical circulatory support devices (MCSD) in Japan : current status and future directions
- Extensive Patch Angioplasty of the Left Main Ostial Stenosis Using a Rhombic-Shaped Pulmonary Autograft
- IS-1-4 100% 1 Year Survival with Complete Pulsatile High Flow Circulatory Support Was Achieved in Japanese Feasibility Study of Implantable Centrifugal LVAD "EVAHEART"
- Long-Term Circulatory Support With a Left Ventricular Assist Device Therapy in Japan
- Cardiac valvular surgery in dialysis patients : comparison of surgical outcome for mechanical versus bioprosthetic valves
- Evaluation of Mechanical Seals for Rotary Blood Pumps by a Chronic Animal Experiment
- Perspectives of Implantable Ventricular Assist Devices
- EVAHEART^ : An Implantable Centrifugal Blood Pump for Long-Term Circulatory Support
- The Effect of Nonlinear Viscoelasticity of Lubricants of Piston Ring Lubrication
- Fluid Film Lubrication with Nonlinear Four-Element Viscoelastic Fluids : Finite-Width Journal Bearings and Parallel Circular Bearings in Normal Approach
- Fluid Film Lubrication with Nonlinear Four-Element Viscoelastic Fluids : Flow Characteristics, a Modified Reynolds Equation and its Application to Parallel Circular Squeeze Bearings : Vibration, Control Engineering, Engineering for Industry
- Evaluation of Mechanical Seals for Rotary Blood Pumps by a Chronic Animal Experiment
- A novel counterpulse drive mode of continuous-flow left ventricular assist devices can minimize intracircuit backward flow during pump weaning
- What is the ideal off-test trial for continuous-flow ventricular-assist-device explantation? Intracircuit back-flow analysis in a mock circulation model
- A novel counterpulsation mode of rotary left ventricular assist devices can enhance myocardial perfusion
- Neointima-inducing inflow cannula with titanium mesh for left ventricular assist device
- Alteration of LV end-diastolic volume by controlling the power of the continuous-flow LVAD, so it is synchronized with cardiac beat : development of a native heart load control system (NHLCS)