Mechanical impedance of the human mastoid and forehead. a critique of the mechanical coupler of the IEC373.:A CRITIQUE OF THE MECHANICAL COUPLER OF THE IEC373
スポンサーリンク
概要
- 論文の詳細を見る
Mechanical impedances of human mastoids and foreheads were measured on frequencies from 250 to 6300Hz for 80 young adults by means of B&K impedance head (# 8000) and shaker (# 4810) as main parts of the measuring setup. The measurements revealed the following problems in the IEC373 (1971/1981) which had defined the mechanical impedance characteristics of the mechanical coupler for calibration of bone vibrators in audiometric use. The average impedance levels of the mastoids differ significantly from those of the foreheads in middle and high frequencies. The impedance levels of the IEC373 mechanical coupler are higher than those of the present measurements of the mastoids and the foreheads. The differences exceed the tolerance limits of IEC373 (1981) in every frequency. These findings strongly suggest that the coupler is neither the artificial mastoid nor the artificial forehead but a coupler having a spedified impedance, therefore, disregard for effect of the type of bone vibrator on the Reference Equivalent Threshold Force Level, as seen in IS07566, should be revised and that the RETFL should be determined for an each type of bone vibrator.
- 一般社団法人 日本耳鼻咽喉科学会の論文