Total skin electron beam therapy using an inclinable couch on motorized table and a compensating filter
スポンサーリンク
概要
- 論文の詳細を見る
Total skin electron beam is a specialized technique that involves irradiating the entire skin from the skin surface to only a few millimetres in depth. In the Stanford technique, the patient is in a standing position and six different directional positions are used during treatment. Our technique uses large electron beams in six directions with an inclinable couch on motorized table and a compensating filter was also used to spread the electron beam and move its intensity peak. Dose uniformity measurements were performed using Gafchromic films which indicated that the surface dose was 2.04 ± 0.05 Gy. This technique can ensure the dose reproducibility because the patient is fixed in place using an inclinable couch on a motorized table.
- American Institute of Physicsの論文
American Institute of Physics | 論文
- Influence of film composition in Co2MnSi electrodes on tunnel magnetoresistance characteristics of Co2MnSi/MgO/Co2MnSi magnetic tunnel junctions
- Tunneling anisotropic magnetoresistance in epitaxial CoFe/n-GaAs junctions
- Highly spin-polarized tunneling in fully epitaxial Co2Cr0.6Fe0.4Al/MgO/Co50Fe50 magnetic tunnel junctions with exchange biasing
- Tunnel magnetoresistance in epitaxial magnetic tunnel junctions using full-Heusler alloy Co2MnGe thin film and MgO tunnel barrier
- Structural and magnetic properties of epitaxially grown full-Heusler alloy Co2MnGe thin films deposited using magnetron sputtering