放射線治療におけるCRポータル画像とDRR画像の位置照合プログラムの開発
スポンサーリンク
概要
- 論文の詳細を見る
In this article, the authors propose an image registration program of portal images and digitally reconstructed radiography (DRR) images used as simulation images for external beam radiation therapy planning. First, the center of the radiation field in a portal image taken using a computed radiograhy cassette is matched to the center of the portal image. Then scale points projected on a DRR image and the portal image are deleted, and the portal image with the radiation field is extracted. Registration of the DRR and portal images is performed using mutual information as the registration criterion. It was found that the absolute displacement misregistrations in two directions (x, y) were 1.2±0.7 mm and 0.5 ±0.3 mm, respectively, and rotation disagreement about the z axis 0.3±0.3°. It was concluded the proposed method was applicable to image registration of portal and DRR images in radiation therapy.
- 日本医学物理学会の論文
- 2012-00-00
日本医学物理学会 | 論文
- 大気中ラドン・トロン濃度に及ぼす地質, 地表面の相違
- 放射線治療計画システムと独立したモニタ単位数の検証計算と出力係数
- 群馬県内施設を対象とした頭部CT検査における撮影線量と画質の比較評価
- 呼吸誘導による放射線治療計画用呼吸同期CT撮影
- 平行平板形電離箱(ルース型およびアドバンスト・マーカス型)の校正定数比の実験的評価