A sheep survived for 48 days with the biventricular bypass type total artificial heart.
スポンサーリンク
概要
- 論文の詳細を見る
<p>A biventricular bypass type total artificial heart (BVB-TAH) utilizing two pusher-plate pumps was developed and implanted in a sheep for 48 days with excellent results. A Hall effect sensor was utilized to operate each pump independently with a full stroke at variable rates (VR). With this system, the animal's hemodynamics was kept physiologically, and all metabolic parameters except hemoglobin and hematocrit returned to normal three weeks after implantation. However, signs of infection appeared on the forty-second day, and consequently the animal fell into a state of shock. Even at that time the BVB-TAH maintained circulation by increasing pumping rate automatically. On the forty-eighth day, the animal could not stand and suffered from anuria; the experiment was then terminated after 1,140 h pumping. At autopsy, there was an enlarged heart with an atrophic change, 1,900 ml of pleural effusion, and 3,100ml of ascites fluid. Blood culture taken on the forty-seventh day yielded Acinetobacter calcoaceticus. The BVB-TAH operated in an independent VR mode maintained entire circulation, and has a capability of substituting the native heart function in any situation.</p>