スぺクトラスパン・プラズマ発光分光光度計による温泉中のホウ素の定量
スポンサーリンク
概要
- 論文の詳細を見る
A plasma emission spectrophotometer "Spectraspan" (low power d. c. plasma arc, operating on argon) was evaluated for boron determination in thermal waters. The influence of acids (hydrochloric, nitric and sulfuric acids) and several metallic ions (Na(+). K(+), Mg(2+). Ca(2+)) on emission intensity was studied. There was a linear relationship between emission intensity and boron content from 0 to 500 mg/l. Boron in thermal waters was easily determined by standard addition method. Precision. coefficient of variance and recoveries of known amount of boron added to the sample for 11 repricate analyses were 0.12μg/ml, 2.02%, 95.0-101.7%, respectively. Boron content of 27 thermal waters in the Sanin district was determined. and the highest B content in the sample waters were 8.8 mg/l (Tottori spa) in Tottori Prefecture and 14.6-25.0 mg/l (the thermal springs at the foot of Mt. Sanbe. a non-active quaternary volcano) in Shimane Prefecture. The relationships between B content and water temperature, B content and pH value were not recognized. Significantly positive correlationship was observed between Band Li contents.
- 1979-03-25
論文 | ランダム
- 『吾妻鏡』における観音・補陀落伝承 : 源頼朝と北条泰時を結ぶ
- 深吸気誘発性上室性貧拍症の1例
- 22-29 黒ボク土露地畑における埋設型ライシメータ法を用いた減肥技術の硝酸溶脱低減評価(22.環境保全,2009年度京都大会)
- 27 施肥窒素の形態変化を繰り入れた硝酸溶脱のモデル化(関東支部講演会,2007年度各支部会)
- 窒素施肥法及び紙マルチ移植栽培が水田系外に排出される汚濁物質量に及ぼす影響と評価