Spectrophotometric Determination of Anionic Surfactants in River Water with Cationic AZO Dye by Solvent Extraction- Flow Injection Analysis
スポンサーリンク
概要
- 論文の詳細を見る
Anionic surfactants in water were determined by a spectrophotometric flow injection technique coupled with solvent extraction. The ion associate which formed between an anionic surfactant and an cationic azo dye was extracted into an organic solvent and the absorbance was measured. The carrier was distilled water, and the reagent solution contained an cationic azo dye and sodium sulfate, the pH of which being adjusted to 5 with acetate buffer. A phase separator with a poly(tetrafluoroethylene) porous membrane (0.8μm pore size) was used to separate the organic phase. Six derivatives of cationic azo dyes and several extracting solvents were examined; a pair of 1-methyl-4-(4-diethylaminophenylazo)- pyridinium cation and chloroform turned out best. The sampling rate was 30 samples per hour. Calibration graphs were linear up to 2×10(-6)M or 3×10(-5)M of anionic surfactant when injection volume was 300 or 100μl, respectively. The relative standard deviation(n=10) was 1.5% for 300μl of 1×10(-6)M sodium dodecylsulfate. The detection limit was as little as 1×10(-8)M of anionic surfactant. Anionic surfactants in river water were determined satisfactorily.
- 日本分析化学会の論文
- 1987-06-10
日本分析化学会 | 論文
- 大気中の粒子状物質の分析
- 機器分析法の基礎知識--X線回折分析法
- 大気粉じんの状態分析
- 化学的アプローチによるナノ領域の操作 : 固体表面における原子, 分子の構造制御集積による機能性物質相の創成
- 加速器質量分析法