Tooth Mobility Measurement of Dental Implants
スポンサーリンク
概要
- 論文の詳細を見る
The use of dental implants has increased together with increases in the human life span and it has become an imperative subject for dentists to familiarize themselves with this treatment modality. Unfortunately, there has been no practical and quantitative method for in vivo evaluation of the stability of dental implants. In the tooth mobility examination, the tactile sense of natural teeth is different from that of dental implants. The authors have developed an automatic diagnosis system of tooth mobility for clinical use. The biomechanical mobility of peri-implantium is measured with a pseudo-random vibration, from which the viscoelasticity c1, c2, k of peri-implantium is obtained. The diagnosis system has been applied to the quantitative evaluation of the stability of implants : endodontics endosseous implants (titanium pin), endosseous implants (Bioceram). It has also been applied to the evaluation of the long-term prognoses of dental implantation (Bioceram) and the examination of Intramobile implant (IMZ), and the satisfactory results have been obtained.
- Faculty of Engineering, Okayama Universityの論文
- 1993-03-25
Faculty of Engineering, Okayama University | 論文