High-sensitive analysis of DNA fragments by capillary gel electrophoresis using transient isotachophoresis preconcentration and fluorescence detection
スポンサーリンク
概要
- 論文の詳細を見る
In this report aimed on further development of a high-sensitivity capillary gel electrophoresis (CCE) method for analysis of DNA fragments, we firstly explored online transient isotachophoresis (tITP) preconcentration combined with fluorescence detection (FD). The fluorescence signal (excitation: 488 nm; emission: 590 nm) was generated using the intercalating dye of ethidium bromide (EB). It was found when the leading electrolyte (LE) was injected behind the sample zone, such a special tITP mode has significant advantages to solve the bubble formation issue and to improve the analytical performance stability. Two standard DNA samples, a 50 bp DNA step ladder and the phi X174/HaeIII digest. were used to evaluate the qualitative and quantitative abilities of the tITP-FD approach. A highly diluted sample (10,000-fold in the water, e.g. the phi X174/HaeIII digest diluted from 500 mu g/ml to the 50 ng/ml level) was enriched and detected; the LOD was down to 0.09 ng/ml for the 72 bp fragment, apparently improved more than 1000-fold in comparison with UV detection. Although the RSD of peak areas (n = 3) was around 15.5% for the sample was electrokinetically injected, good linearity of peak area response showed that the proposed method is suitable for quantitative analysis.