Neutron Dose Rate in the Facility at the Cyclotron Center of Chung Shan Medical University
スポンサーリンク
概要
- 論文の詳細を見る
The neutron dose equivalent rate (DR) leaking from the self-shielded cyclotron was measured using an FHT-751 neutron counting system in the facility at the Cyclotron Center of Chung Shan Medical University. This system was calibrated using two <252>^Cf neutron sources and simulated according to MCNP code. The results show various DRs up to 120 μSv h^<-1> in the cyclotron room. Two-dimensional distributions of measured neutron DRs indicate an explicit, heavy leakage of neutrons through the self-shielded interlock. The neutron DR of the operating cyclotron that is hazardous to the health of medical personal and the public is evaluated.
- 日本放射線影響学会の論文
- 2008-03-16
著者
-
Chen Chien-yi
School Of Medical Imaging And Radiological Sciences Chung Shan Medical Univ.
-
Chen Chien-yi
School Of Medical Imaging And Radiological Sciences Chung Shan Medical University
-
LEE Jui-Ping
Institute of Biochemistry