110 Vibration measurement of MRI device(O.S.4 Measurement and Sensor II)
スポンサーリンク
概要
- 論文の詳細を見る
The MRI inspection is a necessary, indispensable inspection on the medical treatment site of today. A small, piezo-electric obstruction was made to stick to the surface of the device to acquire the vibration characteristic the MRI device's operating and the vibration acceleration was measured. The size and the frequency response of the vibration acceleration generated while operating the MRI device were obtained. The vibration acceleration Z axially reached a value that was larger than Y axially in the center part of the gantry and the opening. The value that rises 100dB in the high frequency element is indicated though the frequency response is a low frequency weighted acceleration level for whole body vibration in the low frequency element. The vibration acceleration the MRI device's operating showed the peak up to the frequency from 500 to 2000Hz. The vibration generation under the inspection cannot be disregarded because of making of the MRI device in recent years high magnetic. It is thought that it wants to reflect it in the control of the device vibration in consideration of the characteristic of the vibration acceleration in the future.
- 一般社団法人日本機械学会の論文
- 2005-05-31
著者
-
Yagi Kazuo
Department Of Health Science Tokyo Metropolitan University Of Health Science University
-
Eguchi Kentaro
Department Of Health Science Tokyo Metropolitan University Of Health Science University
-
Muto Kenji
Tokyo Metropolitan college of Industrial Technology
-
MUTO Kenji
Department of Electronic Engineering, Tokyo Metropolitan College of Aeronautical Engineering
関連論文
- Transmission of Holographic 3D moving picture employed Infrared Network(INTERNATIONAL Workshop on Advanced Image Technology 2008)
- Transmission of Holograohic 3D images using Infrared transmitter(International Workshop on Advanced Image Technology 2007)
- Transmission of Holograohic 3D images using Infrared transmitter
- Mechanical Evaluation of Cardiac Contractility in Left Ventricle with Disease Using Magnetic Resonance Tagging Technique
- Two-Dimensional Deformation Analysis of Human Left Ventricle during Ejection Period Using Magnetic Resonance Tagging Technique
- Data Compression for Transmission of Holographic 3D Images Applying Digital SSTV Processing(International Workshop on Advanced Image Technology 2006)
- Measurement result of slice positioning sound of MRI equipment
- 227 ANALYSIS OF INFLUENCE THAT STRONG MAGNETIC FIELD ENVIRONMENT CAUSES TO NOISE MEASUREMENT SYSTEM(General Session IV)
- q-space Imaging (QSI) of the Brain : Comparison of Displacement Parameters by QSI and DWI
- 110 Vibration measurement of MRI device(O.S.4 Measurement and Sensor II)
- Blind preprocessing method for multichannel feedforward active noise control
- An Investigation on the Plant Modeling Filter's Parameters for Active Noise Control System(Noise and Vibration)
- A-10-12 The Effects of Paths Characteristics on Modeling for Active Control
- A preprocessing method for multichannel feedforward active noise control
- A-10-1 A study on a preprocessing method for multichannel filtered-x algorithm(A-10. 応用音響, 基礎・境界)