The purpose of this study was to investigate the application of nondestructive testing in cyclic fatigue evaluation of endodontic ProFile^[○!R] nickel-titanium (NiTi) rotary instruments. As-received ProFile^[○!R] instruments were made to rotate freely in sloped metal blocks by a contra-angle handpiece mounted on a testing machine. Rotation was interrupted periodically, and the instrument removed and engaged onto a device to monitor its stiffness by using two strain gauges in four different directions. This monitoring method has the potential to be developed into a convenient, nondestructive turnkey system that allows in situ assessment of the integrity of NiTi instruments in the clinic. Upon fracture, which was indicated by a change in instrument stiffness, the fractured surface would be examined under a scanning electron microscope. Microscopic evaluation indicated a small area of fatigue fracture with a large area of final ductile fracture, whereby the latter was the major cyclic failure mode. Based on the results of this study, we concluded that a potential nondestructive integrity assessment method for NiTi rotary instruments was developed.
- 日本歯科理工学会の論文
著者
-
Lan Wan-hong
Graduate Institute Of Clinical Dentistry College Of Medicine National Taiwan University And National
-
Li Uei‐ming
Ohio State Univ. Ohio Usa
-
LI Uei-Ming
Graduate Institute of Clinical Dentistry, College of Medicine, National Taiwan University and Nation
-
SHIN Chow-Shing
Department of Mechanical Engineering, National Taiwan University
-
LIN Chun-Pin
Graduate Institute of Clinical Dentistry, College of Medicine, National Taiwan University and Nation
-
Shin Chow-shing
Department Of Mechanical Engineering National Taiwan University
-
Lin Chun-pin
Graduate Institute Of Clinical Dentistry College Of Medicine National Taiwan University And National
関連論文
- Application of Nondestructive Testing in Cyclic Fatigue Evaluation of Endodontic Ni-Ti Rotary Instruments
- In Vitro Study of DP-bioglass Paste for Treatment of Dentin Hypersensitivity
- In Vitro Study of Dentinal Tubule Occlusion with Sol-gel DP-bioglass for Treatment of Dentin Hypersensitivity