A High Pressure Rolling-Ball Viscometer up to 1 GPa
スポンサーリンク
概要
- 論文の詳細を見る
A rolling-ball viscometer has been developed for use under high pressures up to 1 GPa. It is proven that the calibration coefficient of the viscometer depends on its inclination angle and that this dependence varies with the diameter ratio of the ball and tube. In order to permit a simultaneous measurement of the density, the viscometer was improved, allowing for the attachment of a potentiometric displacement sensor. The precision of the density measurement is estimated to be ±0.22%. The viscosity data of a pressure-transmitting liquid, di-(2-ethyl-hexyl) sebacate (Univis P12), is presented in the range of pressures up to 800 MPa at temperatures from 6 to 60℃. The accuracy of a viscosity measurement is estimated to be +0.95%.
- 社団法人応用物理学会の論文
- 1986-07-20
著者
-
IZUCHI Masaya
National Research Laboratory of Metrology
-
NISHIBATA Ken
National Research Laboratory of Metrology
-
Nishibata K
National Research Laboratory Of Metrology
関連論文
- Photoacoustic Images by Rapid Scanning Method
- Photoacoustic Images by Rapid Scanning Method : Photoacoustic Spectroscopy and Ultrasonic Microscopy
- A High Pressure Rolling-Ball Viscometer up to 1 GPa
- Quantitative Photoacoustic Imaging of Biological Tissues : Waves, Optics and Quantum Electronics
- Growth of Ice IV and Equilibrium Curves between Liquid Water, Ice IV, Ice V and Ice VI
- A Dead-Weight Piston Manometer for Use Up to 2 GPa