Characteristics of a Parallel Plate Avalanche Counter with Thin-Film Electrodes
スポンサーリンク
概要
- 論文の詳細を見る
A small parallel plate avalanche counter (PPAC) consisting of a 0.2 μm-thick stretched polypropylene film was fabricated. Its characteristics as a time-sensitive detector were examined with α-particles from an ^<241>Am source, and a time resolution of 150 psec was obtained. The pulse height distribution of the avalanche electrons was observed and the cause of the distribution was studied. The distribution could be explained as a result of two kinds of fluctuation resulting from energy loss and electron avalanche multiplication.
- 社団法人応用物理学会の論文
- 1983-08-20
著者
関連論文
- Characteristics of a Parallel Plate Avalanche Counter with Thin-Film Electrodes
- Determination of the Townsend Primary Ionization Coefficient Using a Parallel Plate Avalanche Counter