COMPARISON OF SIMULATED LIGHT SENSITOMETER AND X-RAY SENSITOMETER FOR SCREEN-FILM SYSTEM IN MEDICAL RADIOLOGY
スポンサーリンク
概要
- 論文の詳細を見る
Sensitometric techniques are used for determining the characteristic curve in medical imaging. In this study intensity-scale (inverse square law, simulated light) methods are used in the determination of radiographic screen-film system characteristic curve. Films which were exposed to simulated light and X-ray were processed in the same automatic processor simultaneously. These curves were overlapped at Net density of 1.0 to compare the curve shapes. The results show there is a good correlation between the density of X-ray sensitometry and that of simulated light sensitometry. The differences of density are within ±0.12 (±10% log exposure). This applies to both blue sensitive and green sensitive systems.