ECG-gated SPECT(^<99m>Tc)の心機能評価 : QGSの収集処理条件の検討
スポンサーリンク
概要
- 論文の詳細を見る
The QGS program is software for highly reproducible evaluation of perfusion in cardiac muscles and cardiac functions, but basic data collection and processing conditions affect the output values of cardiac function analyses. Because these conditions have not been sufficiently evaluated, we determined in this study the optimum cut off, data collection time, and R-R fractions by comparing the outputs calculated from data obtained with a phantom using QGS with the phantom volume. The optimum cut off was 0.41-0.55 cycle/cm in 8 R-R fractions and 0.41-0.52 cycle/cm in 16 R-R fractions. The values calculated using QGS were underestimated by 11.4% compared with the phantom volume. With longer data collection times, statistical errors were smaller. Errors of data collection in 8 fractions over a period of 20 min were about the same as those in 16 fractions over a period of more than 25 min. Comparison between 8 fractions and 16 fractions in EF levels of patients demonstrated that errors were significantly smaller by 3% in 8 fractions than in 16 fractions, and the level was closer to E/G in 16 fractions than in 8 fractions.
- 公益社団法人日本放射線技術学会の論文
- 2000-08-20