Ultrafast Single-Shot Water and Fat Separated Imaging with Magnetic Field Inhomogeneities
スポンサーリンク
概要
- 論文の詳細を見る
Ultrafast MR imaging (e.g., echo-planar imaging) acquires all the data within only several tens of milliseconds. This method, however, is affected by static magnetic field inhomogeneities and chemical shift; therefore, a high degree of field homogeneity and water and fat signal separation are required. However, it is practically impossible to obtain an homogeneous field within a subject even if in vivo shimming has been performed. In this paper, we describe a new ultrafast MR imaging method called Ultrafast Single-shot water and fat Separated Imaging (USSI) and a correction method for field inhomogeneities and chemical shift. The magnetic field distribution within the subject is measured before the scan and used to obtain images without field inhomogeneity distortions. Computer simulation results have shown that USSI and the correction method can obtain water and fat separated images as real and imaginary parts, respectively, of a complex Fourier transform with a single-shot scan. Image quality is maintained in the presence of field inhomogeneities of several ppm similar to those occurring under practical imaging conditions. Limitations of the correction method are also discussed.
- 社団法人電子情報通信学会の論文
- 1994-08-25
著者
-
Satoh K
Research & Development Center Toshiba Corporation
-
Kuhara S
Toshiba Nasu Works Otawara‐shi Jpn
-
Kanayama S
Toshiba R&d Center Kawasaki‐shi Jpn
-
Kanayama Shoichi
Research & Development Center, TOSHIBA CORPORATION
-
Kuhara Shigehide
Research & Development Center, TOSHIBA CORPORATION
-
Satoh Kozo
Research & Development Center, TOSHIBA CORPORATION