水素フレームイオン化検出器中におけるクロロホルムからのヘキサクロロベンゼンなどの生成
スポンサーリンク
概要
- 論文の詳細を見る
As chlorinated hydrocarbons are said to be converted to more stable compounds by combustion, chloroform was injected into a gas chromatograph equipped with a flame ionization detector (FID) to examine the products. The exaust gas from the FID was sampled by a florisil column and the acetone eluate was analyzed by GC with an electron capture detector or a mass spectrometer when factors such as sample size, flow rate of air and hydrogen supplied to the detector were varied. Injection of 2 microliters of chloroform into a FID through a packed column resulted in formation of hexachlorobenzene together with hexachloroethane, hexachlorobutadiene, pentachlorobenzene and octachloronaphthalene. Formation of hexachlorobenzene was suppressed by making the sample size smaller, increasing the flow rate of hydrogen or decreasing the flow rate of air in the detector.
- 社団法人日本分析化学会の論文
- 1987-02-05
著者
関連論文
- ピリジン及びその誘導体との発色反応による水中のトリクロロエチレンのフローインジェクション分析
- 瀬戸内海における有機スズ化合物の分布と起源
- 真空紫外領域のICP発光分析における分光干渉
- 簡易溶媒抽出-超音波ネブライザー法を用いた誘導結合プラズマ発光分析による水中微量金属の同時定量
- 超音波ネブライザーによる誘導結合プラズマへの有機溶媒の送入とその効果
- 分光学会と財政
- マススペクトルデータ処理によるトレーサー実験により生成したクロロベンゼン類中の炭素-13数分布の解析
- 酸化分解/化学発光法に基づく水中トリクロロエチレンの光ファイバーセンサーの試作
- 溶媒抽出/誘導結合プラズマ発光分光分析法による天然水中微量タリウムの定量
- レーザー蛍光法による二, 三の水中有機汚染物質の遠隔光ファイバーセンシング
- フロ-インジェクション-ICP発光分光分析法による排水中のリン酸イオンと全リンの迅速分別定量
- ナフタレン3塩素化合物の異性体分析
- 高分解能質量分析法によるテトラクロロベンゼンとその重水素化物の分別定量
- 緩衝溶液系における水素化物発生-原子吸光法による微量スズの定量
- 水素フレームイオン化検出器中におけるクロロホルムからのヘキサクロロベンゼンなどの生成
- 溶媒抽出-誘導結合プラズマ発光分析法による水中微量ビスマス,タリウム,インジウムの定量
- 廃棄物焼却に伴い生成する有害有機塩素化合物の分析法について
- 異性体組成の比較による底質中ポリ塩素化ナフタレンの起源の推定
- ガスクロマトグラフィー/ヘリウムマイクロ波誘導プラズマ発光分析法によるベリリウム及びクロムのアセチルアセトン錯体の定量