Using a 3-channel telemeter, an electrocardiograph and a data recorder, the author tried to record vectorcardiogram (VCG) continuously be-fore, during and after exercise. Methods Seventy in- and out-patients in the 2nd Department of Internal Medicine were examined in this study. They included 50 persons without cardiopulmonary disease (33 males, 1 7 females) and 20 patients with complete or incomplete right or left bundle branch block (b.b.b.) (18 males, 2 females). The telemeter employed was a 3-channel FM-FM Model TPE-300R of Fukuda Electro Co. The wave frequencies of the main carrier and the subcarriers were 27.12 Mc/s, and 1.7 kc/s, 3.0 kc/s and 5.4 kc/s, respectively. Four 9-volt dry cells were used as the power supply, and the total time constant was above I .5 seconds. The original Frank's lead system was modified by the author so that the electrodes did not interfere the exercise, the Y electrode at the left leg for the original system being shifted to the left epigastric spot lying on the vertical line passing through the point C of the original system, and the earth electrode at the right leg for the original system to the right epigastric region: At first, scalar electrocardiograms (SECGs) were taken at rest with the original Frank's system and the modified Frank's system in all the 70 cases, using the 3-channel telemeter to get the coefficients for coordinate transformation. Then, the SECGs Of the modified Frank's system were recorded on a magnetic tape by setting the output of the 3-channel telemeter for a Sony data recorder through a multipurposed preamplifier PRE 104 of Kokusai Denshi Kogyo Co. These records were performed both in the 18 healthy and in the 5 with b.b.b., every 30 seconds, for 30 seconds, from a few moments be-fore the beginning of the exercise until 5 minutes after its ending. The magnetic tape was set for the vectorcardiograph of Model FVC-3 of Fukuda Medical Electric Co. to get a VCG. The exercise was performed by the about 3-minute walking at an ordinary speed along a flat hospital corridor which was about 250-meter long and the following 3-minute riding on a fixed bicycle at a speed of 30 km/h. For the purpose of calculating the coefficients of coordinate transformation between the original and the modified Frank's systems, the SECGs at rest were reproduced by fitting the output of the vectorcardiograph to a 3-channel electromagnetic oscillograph of Fukuda Electro Co. The exercise VCG was interpreted theoretically by transforming its coordinates into the original Frank's coordinates through the Coordinate Changer divised by us.
- 社団法人日本循環器学会の論文
- 1970-07-20
著者
関連論文
- 調和分析法による心房細動心弾動図の研究 : 第38回 日本循環器学会東北地方会
- 生体におけるBurgerの三角形について : その求め方に関するわれわれの理論とその実際
- 心室肥大Vector心置図の因子分析について : 第35回日本循環器学会東北地方会
- ベクトル心電図所見の心室肥大鑑別のための評価づけならびにその因子分析への応用
- 36)運動時Vector心電図の研究 : 講演会一般演題
- 心肺疾患における肺動脈圧,心弾動図ならびに尿中カテコールアミン排泄量の相互関係について : 第38回 日本循環器学会東北地方会
- 左室肥大の電量的考察 : 第32回日本循環器学会東北地方会
- ベクトル心電図の線型変換(第7報) : 変換の精度の比較検討 : 第36回 日本循環器学会東北地方会
- 肺動脈圧に及ぼす交感神経β.受容体遮断剤Propranololの影響について : 第34回日本循環器学会東北地方会
- 脚ブロックにおけるFrank誘導VCGと変法Frank誘導YCGとの間の変換係数について : 第31回日本循環器学会東北地方会
- 体位変換台の試作 : 第30回日本循環器学会東北地方会
- 尿中カテコールアミン排泄量と肺動脈圧との関係 (第II報) : 第29回日本循環器学会東北地方学会総会
- ベクトル心電図における誘導ベクトルの間接的算出法並びにその空間構造に関する研究 : 第28回日本循環器学会東北地方会総会
- 5) Vector心電図の座標変換 (第V報) : 第25回日本循環器学会東北地方学会総会
- 15)運動中の心電図並びにVector心電図について : 第23回日本循環器学会東北地方会総会
- テレメーターによる運動時ヘクトル心電図の記録に関する研究 : その方法と臨床成績
- 3)門脈血栓症の1例 : 第22回日本循環器学会東北地方会総会