Development of a New CdTe Area Radiation Monitor using the G(E) Function Method
スポンサーリンク
概要
- 論文の詳細を見る
We developed a small Area Radiation Monitor (ARM) prototype with high accuracy performance by combining a cadmium telluride (CdTe) detector with the G(E) function method. To obtain the measurement accuracy required for ARMs, we created a G(E) function based on detailed spectral responses. The accuracy of the derived dose rate and other expected requirements for practical applications were experimentally evaluated. The accuracy for a CdTe detector with dimensions of 5×5×5 mm3 was within 14% in the 60–300 keV range and within 10% in the 300 keV–1.8 MeV range, which was comparable to a large ARM with an ionization chamber. The accuracy for a CdTe detector with dimensions of 2×2×2 mm3 was within 23% in the 60–300 keV range and within 14% in the 300 keV–1.8 MeV range, which was inferior to the 5×5×5 mm3 detector but higher than a small ARM using a silicon PIN photodiode.
- 社団法人 日本原子力学会の論文
- 2006-10-25
著者
-
内藤 晋
Power And Industrial Systems Research And Development Center Toshiba Corporation
-
NAITO Susumu
Power and Industrial Systems Research and Development Center, Toshiba Corporation
-
IZUMI Mikio
Power and Industrial Systems Research and Development Center, Toshiba Corporation
-
ONODERA Toru
Power and Industrial Systems Research and Development Center, Toshiba Corporation
-
MAEKAWA Tatsuyuki
Power and Industrial Systems Research and Development Center, Toshiba Corporation
-
MORIMOTO Soichiro
Fuchu Complex, Toshiba Corporation
-
OHNO Nobuaki
Toshiba Plant Systems & Services Corporation
-
MITSUBORI Minehisa
Toshiba Plant Systems & Services Corporation
-
Izumi Mikio
Power And Industrial Systems Research And Development Center Toshiba Corporation
-
Maekawa Tatsuyuki
Power And Industrial Systems Research And Development Center Toshiba Corporation
-
Onodera Toru
Power And Industrial Systems Research And Development Center Toshiba Corporation
-
Ohno Nobuaki
Toshiba Plant Systems & Services Corporation
-
Morimoto Soichiro
Fuchu Complex Toshiba Corporation
-
Mitsubori Minehisa
Toshiba Plant Systems & Services Corporation
-
Izumi Mikio
Power And Industrial Systems Res. And Dev. Center Toshiba Corp.
関連論文
- 29pWB-7 レーザー逆コンプトンガンマ線を用いた10〜30MeVにおける^C光分解測定(29pWB 実験核物理領域,理論核物理領域合同 宇宙核物理,実験核物理領域)
- Ion Current Prediction Model Considering Columnar Recombination in Alpha Radioactivity Measurement Using Ionized Air Transportation
- Development of a New CdTe Area Radiation Monitor using the G(E) Function Method
- 電離放射線のイオン流体移送型計測に関する技術開発
- α線の電離空気輸送型計測におけるイオン電流予測モデルの開発
- 24aWF-7 低エネルギーにおける軽核の光反応と少数多体問題
- 23aYC-2 蓄積リングTERASにおける汎用LCSラインの建設
- 26pSB-4 紫外線レーザーを用いたHe ガスTPCの較正
- 31a-YH-2 ターゲットガスTPCによるE1巨大共鳴領域の^4He(γ, pn)d反応の測定
- 29p-YJ-5 ^O(n, γ)^0反応における干渉効果
- 29p-YJ-4 , ^Ne(n, γ)反応と元素合成(2)
- 29p-YJ-3 逆反応を用いた^C(α, γ)^O天体核反応の測定
- 5a-J-5 KENSでの^6Li(n,γ)^7Liの反応断面積測定実験
- 1p-G-13 液体 Nc(n,γ)反応と元素合成(I)
- 1p-G-12 KENS での keV 中性子捕獲実験テスト2
- 21p-D-11 KENSでのkeV中性子捕獲実験テスト1
- 20a-D-10 keV中性子捕獲実験用NaI(Tl)検出器の改良
- 29a-YT-10 ^7Li(n.γ)^8Li反応と^8B太陽ニュートリノ問題
- CdTe放射線検出器を用いた計測技術とその応用