A New Correction Technique for Mass Interferences by ^<40>Ar^<++> and CO_2^<++> during Isotope Analysis of a Small Amount of Ne
スポンサーリンク
概要
- 論文の詳細を見る
To determine the Ne isotopic compositions for small samples with extremely small contents of Ne, mass interferences due to double charge 40Ar and carbon dioxide have to be corrected. Since mass interference due to 40Ar++ significantly affects 20Ne/22Ne ratios, the accurate estimation for the contribution of 40Ar++ is required in Ne analysis. Accurate 40Ar++/40Ar+ ratio (α) is determined from two experiments. First, the peak heights of mass 20 and 40 are measured alternately in a static condition. Second, the same experiment is performed while a trap made of sintered stainless steel is cooled at liquid nitrogen temperature. Since peak heights are proportional to time, four increase rates of peak heights are obtained and α can be calculated from the rates.
- 2004-08-01
著者
関連論文
- Noble gas measurement in individual micrometeorites using laser gas-extraction system
- General characterization of Antarctic micrometeorites collected by the 39th Japanese Antarctic Research Expedition: Consortium studies of JARE AMMs (III)
- A New Correction Technique for Mass Interferences by ^Ar^ and CO_2^ during Isotope Analysis of a Small Amount of Ne
- On low noble gas concentrations in Antarctic micrometeorites collected from Kuwagata Nunatak in the Yamato Meteorite Ice Field
- Report on noble gas isotopic compositions of the 1E and 2E samples: An example of noble gas analyses for small sample sizes
- Remnant extraterrestrial noble gases in Antarctic cosmic spherules
- Mid-infrared transmission spectra of individual Antarctic micrometeorites and carbonaceous chondrites